| Rated capacities | 10kgf |
| 20kgf |
| 50kgf |
| 100kgf |
| 200kgf |
| 300kgf |
| Rated output | 2mV/V±0.2% |
| Maximum safe overload | 200% of rated capacity |
| Combined error | ± 0.05% of rated output |
| Creep | ± 0.1 % of rated capacity/hour |
| Zero balance | 1% of rated capacity |
| Compensated temperature range | -10°C~+60°C/+14°F~+140°F |
| Recommended excitation voltage | 12V |
| Maximum excitation voltage | 15V |
| Input terminal resistance | Approximately 400Ω |
| Output terminal resistance | 350Ω±3.5Ω |
| Insulation resistance | Greater than 5,000MO at DC50V |
| Temperature effect Zero | 0.007% rated output/°C - |
| Span | 0.002% rated output/°CTyp. |
| Cable length | 5rn/195 inches |