Features
- Automatic operation assuring accuracy for everyone at all times
- Stress-minimized, pain-free operation using the forefinger
- Easy calibration (adjustment) by yourself – User CAL function*1*2
- Calibration and display by either volume (µL/mL) or weight (mg)*1
- Accurate pipetting possible with other manufacturers’ tips (List of compatible tips (PDF 40KB))
- Dedicated glass tips and silicon adapters available (Flyer (PDF 336KB))
- Multiple dispensing (MD) mode with automatic initial volume correction (pre-dispensing)
- Mixing (MIX) mode for homogenizing solutions
- Reverse operation for accurately dispensing viscous samples
- Increased resistance to impacts from falls with padding on the head*1
- Aspirating/dispensing speeds adjustable to 5 levels
- Storage for up to 9 preferred settings
- Quickly detachable, autoclavable lower part (121 deg C, 2 atm, 20 minutes)
- Adjustable tip ejector reach
- Long finger hook
- Power supply by either a rechargeable battery or switching (100-240V) AC adapter
- Multiple AC adapter plugs (A, BF, C, S)
- Auto power OFF function
*1 Patent pending
*2 For verification of dispensed volumes necessary for calibration, the AD-4212B-PT / AD-4212A-PT / FX-300iPT pipette accuracy testers (or BM series + BM-014) are useful tools
Specifications
| | | | | | | |
|---|
| Model | | MPA-10 | | MPA-20 | | MPA-200 | |
| Capacity range | | 0.5 to 10.0 µL | | 2.0 to 20.0 µL | | 10 to 200 µL | |
| Increment | | 0.1 µL | | | | 1 µL | |
| Performance*5*6 | Volume | 1.0 µL | 10.0 µL | 2.0 µL | 20.0 µL | 10 µL | 200 µL |
| Accuracy (±) | 4.0% | 1.0% | 4.0% | 1.0% | 2.5% | 0.6% |
| Repeatability (CV) | 2.5% | 0.4% | 2.5% | 0.4% | 1.0% | 0.15% |
| Program memory | | 9 programs | | | | | |
| Aspirating/dispensing speed | | 5 speeds | | | | | |
| Length | | Approx. 280 mm | | | | | |
| Weight (including the battery) | | Approx. 150 g | | | | Approx. 160 g | |
| | | | | | |
|---|
| Model | | MPA-1200 | | MPA-10000 | | |
| Capacity range | | 100 to 1200 µL | | 0.1 to 10.0 mL | | |
| Increment | | 1 µL | | 0.01 mL | | |
| Performance*5 | Volume | 100 µL | 1200 µL | 1.0 mL | 5.0 mL | 10.0 mL |
| Accuracy (±) | 2.5% | 0.5% | 2.5% | 1.0% | 0.5% |
| Repeatability (CV) | 0.6% | 0.15% | 0.6% | 0.15% | 0.15% |
| Program memory | | 9 programs | | | | |
| Aspirating/dispensing speed | | 5 speeds | | | | |
| Length | | Approx. 280 mm | | | | |
| Weight (including the battery) | | Approx. 170 g | | Approx. 190 g | | |
*5 In standard mode with maximum aspirating and dispensing speeds
*6 For the MPA-10, specifications apply for volumes of 1.0 μL or above